5 set up, 1 indication of the switching status, 2 simulation – VEGA VEGASWING S 51 User Manual
Page 18: 3 function chart, 5 s et up
5 S
et up
5
.1
Indication of the switching status
The switching status of the electronics can be
checked via the signal lamp integrated in the upper
part of the housing.
5
.2
Simulation
The VEGASWING S 51 has an integrated function
for simulation of the output signal which can be
activated magnetically. Please proceed as follows:
à
Hold the test magnet (accessory) against the
circle symbol with the label "TEST" on the
instrument housing
Fig. 19: Simulation of the output signal
The test magnet changes the current switching
condition of the instrument. You can check the
change on the signal lamp. Please note that the
connected instruments are activated during the
simulation.
Caution:
It is absolutely necessary that you remove
the test magnet after the simulation from the
instrument housing.
5
.3
Function chart
The following chart provides an overview of the
switching conditions depending on the adjusted
mode and level.
Level
Switching
status
Control
lamp
Mode max.
closed
Green
Mode max.
open
Red
Mode min.
closed
Green
Mode min.
open
Red
Malfunction
any
open
fl
ashes red
18
VEGASWING S 51
Set up
32947
-01
-120124