beautypg.com

VEGA VEGAFLEX 66 (-110…+400°C) Profibus PA User Manual

Page 49

background image

Output variable

Output signal

digital output signal, format according to
IEEE-

754

Sensor address

126 (default setting)

Current value

10 mA, ±0.5 mA

Integration time (63 % of the input
variable)

0 … 999 s, adjustable

Step response or adjustment time

250 ms (ti: 0 s, 10 … 90 %)

Accuracy (similar to DIN EN 60770-1)

Reference conditions according to DIN EN 61298-1

-

Temperature

+

18 … +30 °C (+64 … +86 °F)

-

Relative humidity

45 … 75 %

-

Air pressure

860 … 1060 mbar/86 … 106 kPa
(

12.5 … 15.4 psig)

Deviation in characteristics and characteristics

Reference installation conditions

-

Flange size

> DN

100

-

min. distance to installations

>

500 mm (19.69 in)

-

Vessel diameter

1 m (3.281 ft) centrical installaton

-

Medium

Water

Temperature drift (current output)

0.06 %/10 K relating to the max. measuring
range

Accuracy

see diagrams

3 mm

(

1

/

8

")

-3 mm

(-

1

/

8

")

10 mm

(

25

/

64

")

-10 mm

(-

25

/

64

")

0,5 m

(1.6 ft)

0,08 m

(0.26 ft)

6 m

(20 ft)

Fig. 21: Accuracy - Rod version -110 … +400 °C (-166 … +752 °F)

VEGAFLEX

66 • Profibus PA

49

10 Supplement

32764

-

EN

-080716