6.3 Adjustment system

1.1

2

3

1

Fig. 17: Indicating and adjustment elements
1

LC display

2

Indication of the menu item number

3

Adjustment keys

l

[OK]

key:

-

Move to the menu overview

-

Confirm selected menu

-

Edit parameter

-

Save value

l

[->]

key to select:

-

menu change

-

list entry

-

Select editing position

l

[+]

key:

-

Change value of the parameter

l

[ESC]

key:

-

interrupt input

-

jump to the next higher menu

The sensor is adjusted via the four keys of the indicating and
adjustment module. The LC display indicates the individual
menu items. The functions of the individual keys are shown in
the above illustration. Approx. 10 minutes after the last
pressing of a key, an automatic reset to measured value
indication is triggered. Any values not confirmed with [OK] will
not be saved.