3 exchanging the electronics module – VEGA VEGAFLEX 66 (-110…+400°C) Profibus PA User Manual
Page 41

?
E
017
l
Adjustment span too small
à Carry out a fresh adjustment and increase the distance
between min. and max. adjustment
?
E
036
l
no operable sensor software
à Carry out a software update or send the instrument for
repair
?
E
042/E043
l
Hardware error, electronics defective
à Exchange instrument or return instrument for repair
?
E
113
l
Communication conflict
à Exchange instrument or return instrument for repair
Depending on the failure reason and measures taken, the
steps described in chapter "Set up" must be carried out again,
if necessary.
8.3 Exchanging the electronics module
If the electronics module is defective, it can be replaced by the
user.
In Ex applications only one oscillator with respective Ex
approval may be used.
If there is no electronics module available on site, one can be
ordered from the agency serving you.
The new oscillator must contain the order data of the sensor.
These can be loaded as follows:
l
At the factory by VEGA
l
Or on site by the user
Information:
When loading on site, first of all the order data must be
downloaded from the Internet (see operating instructions
manual "Oscillator").
Reaction after fault rec-
tification
Preparations
VEGAFLEX
66 • Profibus PA
41
8 Maintenance and fault rectification
32764
-
EN
-080716