beautypg.com

VEGA VEGAFLEX 66 (-20…+250°C) Foundation Fieldbus User Manual

Page 52

background image

Voltage supply

Operating voltage

-

Non-Ex instrument

9 … 32 V DC

-

EEx-ia instrument

9 … 24 V DC

-

EEx-d instrument

16 … 32 V DC

Operating voltage with lighted indicating and adjustment module

-

Non-Ex instrument

12 … 32 V DC

-

EEx-ia instrument

12 … 24 V DC

-

EEx-d instrument

20 … 32 V DC

Power supply by/max. number of sensors

-

H

1 power supply

max. 32 (max. 10 with Ex)

Electrical protective measures

Protection, depending on housing version

-

Plastic housing

IP

66/IP 67

-

Aluminium housing, stainless steel
housing - investment casting, stainless
steel housing - electro-polished

IP

66/IP 68 (0.2 bar)

5)

-

Aluminium and stainless housing, in-
vestment casting (optionally available)

IP

66/IP 68 (1 bar)

Overvoltage category

III

Protection class

II

Approvals

Depending on the version, instruments with approvals can have different technical data.

For these instruments, the corresponding approval documents have to be taken into account.
These are part of the delivery or can be downloaded under www.vega.com via "VEGA Tools" and
"

serial number search" as well as via "Downloads" and "Approvals".

5)

A suitable cable is the prerequisite for maintaining the protection rating.