Process conditions

Process pressure

-1 … +100 bar/-100 … +10000 kPa (-
-14.5 … +1450 psig), depending on the process
fitting

Process temperature

-20 … +250 °C (-4 … +482 °F)

The measurement error from the process condi-
tions is in the specified pressure and temperature
range of below 1 %.

1

2

0°C

(32°F)

-20°C

(-4°F)

0°C

(32°F)

40°C

(104°F)

20°C

(68°F)

70°C

(158°F)

80°C

(176°F)

50°C

(122°F)

150°C

(302°F)

100°C

(212°F)

250°C

(482°F)

200°C

(392°F)

Fig. 26: Ambient temperature - process temperature (version -20 … +250 °C/-4 … +482 °F)
1

Ambient temperature

2

Process temperature (depending on the seal material)

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar

Cable entry/plug

4)

-

Single chamber housing

l

1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5

or:

l

1 x closing cap M20 x 1.5; 1 x blind stopper
M

20 x 1.5

or:

l

1 x closing cap ½ NPT, 1 x blind plug ½ NPT

or:

l

1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5

-

Double chamber housing

l

1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5; 1 x blind stopper
M

16 x 1.5 or optionally available with 1 x plug

M

12 x 1 for VEGADIS 61

or:

4)

Depending on the version M12 x 1, according to DIN 43650, Harting, 7/
8" FF.

50

VEGAFLEX

66 • Foundation Fieldbus

10 Supplement

34172

-EN

-100426