beautypg.com

VEGA VEGABAR 63 Foundation Fieldbus User Manual

Page 40

background image

This function enables reading out parameter adjustment data as well
as writing parameter adjustment data into the sensor via the indicating
and adjustment module. A description of the function is available in the
operating instructions manual "Indicating and adjustment module".

The following data are read out or written with this function:

l

Measured value presentation

l

Adjustment

l

Damping

l

Linearisation curve

l

Sensor-TAG

l

Displayed value

l

Unit of measurement

l

Language

The following safety-relevant data are not read out or written:

l

PIN

l

Application

Copy sensor data

Copy sensor data?

Basic adjustment
If the "Reset" is carried out, the sensor resets the values of the
following menu items to the reset values (see chart):

8)

Menu section

Function

Reset value

Basic settings

Unit of measurement

bar

Temperature unit

°C

Zero/Min. adjustment

Measuring range begin

Span/Max. adjustment

Measuring range end

Density

1 kg/l

Density unit

kg/l

Damping

0 s

Linearisation

linear

Sensor-TAG

Sensor

Display

Displayed value

AI-Out

The values of the following menu items are not reset with "Reset:

Menu section

Function

Reset value

Basic settings

Position correction

no reset

Display

Lighting

no reset

Service

Language

no reset

8)

Sensor-specific basic adjustment.

Copy sensor data

Reset

40

VEGABAR

63 • Foundation Fieldbus

6 Set up with the indicating and adjustment module PLICSCOM

32464

-EN

-081209