4 Mounting

VEGACAL 67 • Profibus PA

31759-EN-130917

4 Mounting

4.1 General instructions

Make sure that all parts of the instrument coming in direct contact

with the process, especially the sensor element, process seal and

process fitting, are suitable for the existing process conditions, such

as process pressure, process temperature as well as the chemical

properties of the medium.
You can find the specifications in chapter "Technical data" and on the

nameplate.

Select an installation position you can easily reach for mounting and

connecting as well as later retrofitting of a display and adjustment

module. The housing can be rotated by 330° without the use of any

tools. You can also install the display and adjustment module in four

different positions (each displaced by 90°).

Before beginning the welding work, remove the electronics module

from the sensor. By doing this, you avoid damage to the electronics

through inductive coupling.

With threaded versions, the housing must not be used to screw in the

instrument. Applying tightening forces on the housing can damage its

internal parts.
Use the hexagon for screwing in.

Use the recommended cables (see chapter "Connecting to power

supply") and tighten the cable gland.
You can give your instrument additional protection against moisture

penetration by leading the connection cable downward in front of the

cable entry. Rain and condensation water can thus drain off. This ap-

plies mainly to outdoor mounting as well as installation in areas where

high humidity is expected (e.g. through cleaning processes) or on

cooled or heated vessels.

Fig. 4: Measures against moisture penetration

The process fitting must be sealed if there is gauge or low pressure in

the vessel. Before use, check if the seal material is resistant against

the measured product and the process temperature.

Suitability for the process

conditions

Installation position

Welding work

Handling

Moisture