beautypg.com

VEGA VEGAWELL S 51 User Manual

Page 23

background image

Reference conditions and actuating variables (similar to DIN EN 60770-1)

Reference conditions according to DIN EN 61298-1

-

Temperature

+15

… +25 °C

-

Relative humidity

45 … 75 %

-

Air pressure

860 … 1060

mbar/86 … 106 kPa

Determination of characteristics

Limit point adjustment according to IEC 61298-2

Characterstic curve

Linear

Reference installation position

upright, diaphragm points downward

Influence of the installation position

< 0

.2

mbar/20 Pa

Deviation determined according to the limit point method according to IEC 60770

5

Deviation

< 0

.2 %

Influence of the product or ambient temperature

Average temperature coefficient of the zero signal

< 0

.15 %/10 K

Thermal change, current output

< 0

.15 %

with -40 … +60 °C

Long-term stability (similar to DIN 16086, DINV 19259-1 and IEC 60770-1)

Long-term drift of the zero signal

< 0

.1 % /2

years

Ambient conditions

Ambient temperature

-40 … +60 °C

Storage and transport temperature

-40 … +100 °C

Process conditions

Product temperature

-20 … +60 °C

Vibration resistance

mechanical vibrations with 4 g and 5 … 100 Hz

Electromechanical data

Suspension cable

-

Structure

six wires, one suspension cable, one breather
capillary, screen braiding, foil, mantle

-

Wire cross-section

0.5

mm²

-

Wire resistance

≤ 0

.036 O

hm/m

-

Tensile strength

≥ 1200 N

-

Max. length

1000

m

-

Min. bending radius

25

mm

-

Diameter

approx. 8 mm

Voltage supply

Supply voltage

9.6 … 36 V DC

VEGAWELL S 51

23

Supplement

32946

-

01

-

101123