5 software update – VEGA VEGAFLEX 62 Foundation Fieldbus User Manual
Page 42
Information:
When loading on site, first of all the order data must be downloaded
from the Internet (see operating instructions manual "Oscillator").
The electronics modules are adapted to the respective sensor and
distinguish also in the signal output or power supply.
Electronics module FX-E.60F suitable for VEGAFLEX 61, 62, 63, 65,
66 - Foundation Fieldbus:
l
FX-E.60FX (X = without approvals)
l
FX-E.60FC (C = approvals CX, CK, GX according to product list)
8
.5 Software update
The software version of VEGAFLEX 62 can be determined as follows:
l
on the type label of the electronics
l
via the indicating and adjustment module
l
via PACTware
You can view all software histories on our website www.vega.com.
Make use of this advantage and get registered for update information
via e-mail.
The following components are required to update the sensor software:
l
Sensor
l
Voltage supply
l
VEGACONNECT
l
PC with PACTware
l
Current sensor software as file
At "www.vega.com/downloads" go to "Software". Select under "plics
instruments and sensors" the suitable instrument series. Load the zip
file via the right mouse key with "Save target as" e.g. on the desktop of
your PC. Extract all files available in the zip file, e.g. to the desktop.
Connect the sensor to power supply and provide connection from PC
to the instrument via VEGACONNECT. Start PACTware and provide
connection to the sensor, e.g. via the VEGA project assistant. Close
the parameter window of the sensor, as far as open.
Go in the PACTware menu bar to "Instrument data", "Additional
functions" and "Update instrument software".
PACTware now checks the actual hardware and software version of
the sensor and displays the data. This procedure lasts approx. 60 s.
Push the button "Update software" and select the previously
extracted hex file. Then the software update can be started. The
additional files are installed automatically. Depending on the sensor,
this procedure can last approximately 1 h.
Assignment
Foundation Fieldbus
Load sensor software to
PC
Prepare update
Load software into sen-
sor
42
VEGAFLEX
62 • Foundation Fieldbus
8 Maintenance and fault rectification
31840
-EN
-100426