.2 Principle of operation

VEGAWELL

51

is used for level measurement in vessels and

for gauge measurement in deep wells, e.g. in water/sewage
water applications.

1)

T

he actual sensor element is the CERTEC

®

measuring cell

with rugged ceramic diaphragm. The hydrostatic pressure
causes a capacitance change in the measuring cell via the
ceramic diaphragm. This change is converted into an
appropriate output signal.

T

he CERTEC

®

measuring cell is also equipped with a

temperature sensor PT 100 in four-wire technology. The
resistance value can be gripped via the wires of the
suspension cable. The power supply or the processing is
carried out via a temperature sensor, e.g. in the VEGABOX 02.

T

wo-wire electronics 4 … 20 mA/HART for power supply and

measured value transmission over the same cable.

T

he supply voltage range can differ depending on the

instrument version.

D

ata for power supply are specified in chapter "Technical

data".

3

.3 Operation

VEGAWELL

51 - 4 … 20

mA/HART can be adjusted with

different adjustment media:

l

an adjustment software according to FDT/DTM standard,
e.g. PACTware™ and PC

l

W

ith a HART handheld

T

he kind of adjustment and the adjustment options depend on

the selected adjustment component. The entered parameters
are generally saved in the respecitive sensor, when adjusting
with PACTware™ and PC optionally also in the PC.

3

.4 Packaging, transport and storage

Y

our instrument was protected by packaging during transport.

I

ts capacity to handle normal loads during transport is assured

by a test according to DIN EN 24180.

1)

F

or use in closed vessels under vacuum, VEGAWELL 51 is available with

absolute pressure measuring ranges.

Application range

Functional principle

Voltage supply

Packaging

10

VEGAWELL

51 • 4 … 20

mA/HART

P

roduct description

32468

-EN

-080604