Insert the transducer tube in the requested length from below into
the swivelling holder

4

Fasten with screws (4); torque max. 10 Nm

5

Remove the plug from below out of the instrument housing and
plug it into the socket of the transducer tube

6

Plug the electronics housing to the transducer tube. Do not
squeeze any cables. The housing is in the correct position, if the
lower notch (2) of the transducer tube is no longer visible

7

Tighten the screws (5) on the housing moderately with an Allen
wrench (size 4)

8

Loosen terminal screw of the swivelling holder (3) with a fork
spanner (SW 13)

9

Direct the sensor with the swivelling holder to the measured
product

10 Fasten the swivelling holder with screw (3), torque max. 20 Nm
11 Secure screw (3) with lacquer or similar to ensure permanent

tightness of the swivelling holder

12 Rotate the housing in such a way that the cable entry points

downwards (moisture can drain off). Fasten the housing with
screws (5); torque Aluminium housing max. 5 Nm, torque plastic
housing max. 2 Nm

1

2

3

4

5

Fig. 5: Mounting - Version B
1

Groove to lock the housing

2

Notch

3

Screw of swivelling holder (hexagon SW 13)

4

Grub screw, fixing ring (hexagon spanner - size 4)

5

Terminal screws, housing (hexagon spanner - size 4)

14

VEGASON

64 • Foundation Fieldbus