beautypg.com

VEGA VEGASON 64 Foundation Fieldbus User Manual

Page 12

background image

The reference plane for the measuring range (version A) is the lower
edge of the flange. The reference plane for versions B, C and D is the
lower edge of the transducer.

Remember that with all versions there is a minimum distance below
the reference plane - the so-called dead band - in which measurement
is not possible; this distance from reference plane to medium must not
be underrun. The exact value of the dead band is stated in chapter
"T

echnical data".

Information:
If the medium reaches the transducer, buildup can form on it and
cause faulty measurements later on.

1

3

2

100%

0%

4

Fig. 3: Version A - Measuring range (operating range) and max. measuring
distance
1

full (dead band)

2

empty (max. measuring distance)

3

Measuring range

4

Reference plane

Measuring range

12

VEGASON

64 • Foundation Fieldbus

4 Mounting

28793

-EN

-081127