VEGA VEGASON 63 Foundation Fieldbus User Manual
Page 43
43
10 Supplement
VEGASON 63 • Foundation Fieldbus
28792-EN-130417
Ʋ Standard length
5 m (16.4 ft)
Ʋ Max. length
1000 m (3280 ft)
Ʋ Min. bending radius
25 mm (0.984 in) with 25 °C (77 °F)
Ʋ Diameter approx.
8 mm (0.315 in)
Ʋ Colour - standard PE
Black
Ʋ Colour - standard PUR
Blue
Ʋ Colour - Ex-version
Blue
Display and adjustment module
Voltage supply and data transmission
through the sensor
Indication
LC display in dot matrix
Adjustment elements
4 keys
Protection rating
Ʋ unassembled
IP 20
Ʋ mounted into the sensor without cover IP 40
Material
Ʋ Housing
ABS
Ʋ Inspection window
Polyester foil
Voltage supply
Operating voltage
Ʋ Non-Ex instrument
9 … 32 V DC
Ʋ Ex-ia instrument
9 … 24 V DC
Ʋ Exd instrument
16 … 32 V DC
Operating voltage with illuminated display and adjustment module
Ʋ Non-Ex instrument
12 … 32 V DC
Ʋ Ex-ia instrument
12 … 24 V DC
Ʋ Exd instrument
20 … 32 V DC
Power supply by/max. number of sensors
Ʋ Fieldbus
max. 32 (max. 10 with Ex)
Electrical protective measures
Protection, depending on housing version
Ʋ Plastic housing
IP 66/IP 67
Ʋ Aluminium housing, stainless steel
housing - investment casting, stain-
less steel housing - electro-polished
IP 66/IP 68 (0.2 bar)
10)
Overvoltage category
III
Protection class
II
Approvals
Instruments with approvals can have different technical data depending on the version.
10)
A suitable cable is the prerequisite for maintaining the protection rating.