beautypg.com

5 mounting steps version b – VEGA VEGASON 66 Emitting electronics User Manual

Page 12

background image

1

2

3

Fig. 11: Mounting the new emitting electronics
1

Emitting electronics

2

Torx screws T15 (4 pcs.)

3

Flange adapter

2

Plug the new emitting electronics onto the flange adapter.

3

Retighten the Torx screws

4

.5 Mounting steps version B

Danger:
Individual wires of the transducer cable have a voltage of
approx. 70 V during operation (see chapter "Mounting
preparations
"). There is danger of electric shocks! Carry out
the mounting steps only in the complete absence of line
voltage.

Proceed as follows:

1

Loosen the terminal screws (2) on the housing with the
hexagon spanner size 4

2

Remove the previous emitting electronics

Dismount the pre-
vious emitting elec-
tronics

12

Emitting electronics - for VEGASON 64, 65, 66

Mounting

30205

-EN