31

Instrument modification

6.2 Exchange of electronics

You first have to remove the adjustment mod-
ule as described in chapter "6.1 Exchange of
adjustment modules“ to exchange the com-
plete electronics module of VEGABAR.

• Then remove the ground screw and the

two smaller fixing screws that hold the
electronics unit to the housing

• Pull the electronic unit to the top and pull

out the plug connection

• Proceed in reverse sequence when install-

ing the new electronics module

Ground
screw

Fixing screw

Fixing screw

Plug
connection

Note:

When switching on the first time (connection
of power supply) after exchange of the elec-
tronics unit, it will take approx. 20 s until the
actual measured value is displayed.