beautypg.com

2 technical data, 1 data – VEGA D98 Pressure transmitter User Manual

Page 4

background image

4

Pressure transmitter D98

Technical data

Power supply and signal circuit

Power supply
- without indication

12 … 36 V DC

Residual ripple of the power supply

no influence with U

SS

1 V

Output signal

4 … 20 mA (factory setting)

Current limitation

approx. 22 mA

Fault signal

approx. 22 mA

Average delay time

20 ms

Max. permissible load

see load diagram:

12

300

600

900

18

24

30

36

load in

supply voltage
in V

2 Technical data

2.1 Data

Nominal measuring range

Gauge pressure resistanLow pressure resistance

Gauge pressure

0…0.4 bar / 0…40 kPa

30 bar / 3 000 kPa

-0.8 bar / -80 kPa

0…1.0 bar / 0…100 kPa

35 bar / 3 500 kPa

-1.0 bar / -100 kPa

0…2.5 bar / 0…250 kPa

50 bar / 5 000 kPa

-1.0 bar / -100 kPa

0…4.0 bar / 0…400 kPa

65 bar / 6 500 kPa

-1.0 bar / -100 kPa

0…10.0 bar / 0…1 000 kPa

90 bar / 9 000 kPa

-1.0 bar / -100 kPa

0…16.0 bar / 0…1 600 kPa

130 bar / 13 000 kPa

-1.0 bar / -100 kPa

Measuring accuracy (according to DIN 16 086, DIN V 19 259-1 and IEC 770)

Measuring deviation

Reference conditions acc. to IEC 770, e.g.
- temperature

15°C … 35°C

- rel. humidity

45 % … 75 %

- air pressure

860 mbar … 1060 mbar

Determination of characteristics

limit point adjustment acc. to DIN 16 086

Characteristics

linear

Deviation in characteristics

1)

< 0.25 %

Hysteresis (repeatability)

1)

< 0.02 %

1)

Relating to the nominal range

2)

In compensated temperature range of 0°C … +80°C, reference temperature 20°C

3)

According to IEC 770, Punkt 6.3.2 related to the nominal range

4)

The use of a properly fitting cable seal in the cable connection is necessary for maintaining the housing
protection IP 66. If the seal used does not fit the cable, it should be replaced by a suitable one.

5)

FDA approved