6 instrument modification, 1 exchange of hygienic form seal – VEGA D98 Pressure transmitter User Manual

Page 11

background image

Pressure transmitter D98

11

1 Pressure transmitter
2 Headless screw (avoids overwinding and must not be

removed)

3 Adapter
4 Hexagon screw
5 Form seal
6 Process connection

Instrument modification

6 Instrument modification

6.1 Exchange of hygienic form seal

In the hygienic connections of pressure
transmitter D98, the ceramic measuring cell is
gaplessly sealed with a radial, form seal.
This form seal (material EPDM-FDA) can be
exchanged by the user without a fresh ad-
justment being necessary. The criteria and
the time intervals for such an exchange of the
seal are generally defined by the process
and hygienic requirements.

Important:

Use only original replacement seals from
VEGA, article no. 2.17 775!

Please note the following exchange proce-
dure.

Removal procedure:

1 Pressure transmitter should be

unpressurized (switch off process pres-
sure or empty vessel)

2 Loosen the hexagon pressure screw (4)

(width across flats, 46 mm) by turning to
the left.

Note!

The process connection (6) (e.g. compres-
sion nut) must not be loosened!

3 Turn the adapter (3) below the pressure

transmitter housing with a spanner (with
across flats, 36 mm) slightly to the left
(loosen the seal).

4 Hold onto the housing (1), pull the entire

pressure transmitter out of the process
connection by carefully turning to the left.

5 Slightly lift the form seal (5) and loosen it

from the ceramic measuring cell.

6 If the form seal (5) does not completely

encircle the measuring cell, the remaining
part must be carefully removed from the
process connection.

Installation procedure:
1 Place the new form seal (5) over the meas-

uring cell (the conical side must mate with
the process connection).

2 Turning it lightly clockwise, insert the pres-

sure transmitter carefully into the process
connection.

3 Tighten hexagon pressure screw (4) (width

across flats, 46 mm) (45 Nm).

4 Turn the housing (1) of the pressure trans-

mitter to the original position.

The exchange of the form seal is now fin-
ished; the removed seal must be properly
disposed of.