Power supply

Operating voltage

-

Non-Ex instrument

9 … 32 V DC

-

EEx-ia instrument

9 … 24 V DC

-

EEx-id instrument

9 … 32 V DC

Operating voltage with lighted indicating and adjustment module

16)

-

Non-Ex instrument

12 … 36 V DC

-

EEx-ia instrument

12 … 30 V DC

-

EEx-id instrument

12 … 32 V DC

Power supply by/max. number of sensors

-

DP/PA segment coupler

max. 32 (max. 10 with Ex)

-

VEGALOG

571 EP card

max. 15 (max. 10 with Ex)

Electrical protective measures

Protection rating

-

Housing, standard

IP

66/IP 67

17)

-

Aluminium and stainless housing (op-
tionally available)

IP

68 (1 bar)

18)

-

Process component in IP 68 version

IP

68

-

External housing

IP

65

Overvoltage category

III

Protection class

II

Approvals

Depending on the version, instruments with approvals can have different technical data.

For these instruments, the corresponding approval documents have to be taken into account.
These are part of the delivery or can be downloaded under www.vega.com via "VEGA Tools" and
"serial number search" as well as via "Downloads" and "Approvals".

16)

Is available at a later date for instruments with StEx, WHG or ship approval
as well as country-specific approvals such as those according to FM or
CSA.

17)

Instruments with gauge pressure measuring ranges cannot detect the am-
bient pressure when submerged, e.g. in water. This can lead to falsification
of the measured value.

18)

Only with instruments with absolute pressure ranges.

60

VEGABAR

65 • Profibus PA

10 Supplement

28253

-EN

-090522