beautypg.com

VEGA VEGABAR 65 Profibus PA User Manual

Page 33

background image

6

Enter the requested density value with [->] and [+], confirm with
[OK]

and move to position correction with [->].

The adjustment unit is hence switched over from bar to m.

Proceed as follows to select the temperature unit:

4)

à Activate the selection with [OK] and select "Temperature unit

with [->].

à Activate the selection with [OK] and select the requested unit with

[->]

(e.g. °F).

à Confirm with [OK].
The temperature unit is hence switched over from °C to °F.

Proceed as follows:

1

Activate in the menu item "Position correction" the selection with
[OK]

.

Position correction
Offset

P

=

+

0000 mbar

53 mbar

2

Select with [->], e.g. to accept actual measured value.

Position correction
Accept current measured

value?

Accept

Edit

3

Confirm with [OK] and move to min.(zero) adjustment with [->].

Proceed as follows:

1

Edit the % value in the menu item "Min. adjustment" with [OK].

Min. adjustment

+000

.0 %

=

+0000

.0 mbar

0000.0 mbar

2

Set the requested percentage value with [+] and [->].

3

Edit the requested mbar value with [OK].

4

Set the requested mbar value with [+] and [->].

5

Confirm with [+] and move to max. adjustment with [->].

The min. adjustment is finished.

4)

Selection options: °C, °F.

Carry out position cor-
rection

Carrying out min. ad-
justment

VEGABAR

65 • Profibus PA

33

6 Set up with the indicating and adjustment module PLICSCOM

28253

-

EN

-090522