beautypg.com

VEGA VEGAFLEX 55K User Manual

Page 9

background image

VEGAFLEX 55K

9

Product description

Accuracy

(under reference conditions acc. to IEC 770 - relating to the max. meas. range)

Linearity error

< 0.1 %

Temperature drift

0.015 %/10 K

Resolution of the 4 … 20 mA signal

0.025 % of range (DA converter)

Resolution

1 mm

Characteristics

Min. span between
full and empty adjustment
- analogue output signal

10 mm

Ambient conditions

Vessel pressure

-1 … 40 bar

Ambient temperature on the housing

-40°C … +60°C

Process temperature

-40°C … +150°C

Storage and transport temperature

-40°C … +80°C

Protection

IP 66/IP 67 (meets both protection standards)

Protection class
- two-wire sensor

II

- four-wire sensor

I

Overvoltage category

III

Min. dielectric constant

ε

r

> 1.5

Process fittings

VEGAFLEX 55

G

3

/

4

A,

3

/

4

" NPT, G 1A, 1" NPT, G 1

1

/

2

A, 1

1

/

2

" NPT

1.4435 (stst) or Hastelloy C22 plated
flanges

Ex technical data

Ex-Zone 0 acc. to ATEX II 1/2 G
Ex Zone 0 EEx ia IIC T 6 acc. to ATEX II 1/2 G
Ex Zone 0 EEx d ia IIC T 6 acc. to ATEX II 1/2 G
The permissible operating data of the VEGAFLEX sensors for Ex areas are listed in the
certificate.

Materials

Housing

PBT (Valox) or
Aluminium (powder-coated)

Coax rod

1.4435 (stst: 316L) or Hastelloy C22