ACU-RITE

Error Messages

If the communication lines between the

VRO 300

and the PC are disrupted, an error

message will appear on the

VRO 300

screen. Settings that do not match and a

faulty serial cable connection are the most common problems and are easiest to
correct.

Error

Cause

Solution

Baud rate error

VRO 300

and PC baud

rate settings do not
match.

Change one of the settings so that
the

VRO 300

and the PC baud rates

are the same.

Checksum error

VRO 300

and PC baud

rate settings do not
match.

Change one of the settings so that
the

VRO 300

and the PC baud rates

are the same.

Baud rate is too slow.

Increase the baud rates on both the

VRO 300

and the PC.

No response
from PC

Serial cable connection
and COM port setting
do not match.

If the COM port setting (1 or 2) on
the PC screen does not correspond
with the port the serial cable is
connected into:
a) Change the setting; or
b) Disconnect the cable from the

COM port that it is connected
to and connect it to a different
COM port.

Incorrect serial cable

Verify that you are using a serial
cable and that it is not a NULL
modem cable. Replace the cable, if
necessary.

VRO 300

and PC baud

rate settings do not
match.

Change one of the settings so that
the

VRO 300

and the PC baud rates

are the same.

The selected COM
port is disabled.

a) Change the PC’s COM port

setting and connect the serial
cable into the appropriate port

b) Exit the remote storage pro-

gram on the PC, then refer to
your Microsoft

®

Windows

®

and PC operator’s manuals for
instructions on how to enable a
COM port.

Timeout error

Loose serial cable

Check that each end of the serial
cable fits snugly into the

VRO 300

serial port and the PC COM port