Activeforever Mettler Sys*Stim 294 4-Channel Neuromuscular Stimulator User Manual

Page 5

background image

5

Microcurrent Mode

+

-

or

+

-

Figure 3.13—Microcurrent

Waveform

Waveform Type:

Monophasic or biphasic

square

wave

Polarity:

Positive or negative or

biphasic

pulses

Volts:

1 Volt peak, 1 Kohm load

Current: 10-990

µA peak, 1 Kohm load

Average current at
maximum intensity
and frequency:

990

µA

Maximum current
density under 2"
diameter electrode.

24.4

µA/cm²

Frequency:

0.5-500

Hz

Duty

Cycle:

50%zHz

Frequency Modulation:

No

Pulse Duration:

1-1000 ms

Available Amplitude
Modulation Options:

Continuous

Available Channels:

Channel Two only

Amplitude Modulation Specifications:

Vector rotation:

Interferential Mode Only

-50%

amplitude

modulation

in

anti phase with an eight second modulation period.

Surge Mode:

Premodulated, Medium Frequency and Biphasic (TNS) Pulsed Modes

Up

ramp:

3

seconds

Down

ramp:

2

seconds

Preset on/off times:

10 seconds on, 10 seconds off
10 seconds on, 20 seconds off
10 seconds on, 30 seconds off
10 seconds on, 40 seconds off
10 seconds on, 50 seconds off
10 seconds on, 60 seconds off

Programmable On time:

1–240 seconds

Programmable Off time:

1–240 seconds

Reciprocation mode:

Premodulated, Medium Frequency and Biphasic (TNS) Pulsed Modes

Up and down ramps:

1 second, reciprocation only

Reciprocation time:

2–240 seconds, (On time = off time)

Combine with Surge:

Use up and down ramps of surge program

Use on/off times of surge program.

Two timer option:

No