Output channel: 2 channels, independently controlled
Pulse Shape:
Symmetric, biphasic rectangular pulse.
(1) TENS & EMS
(200
Ω - 1 kΩ load)
f:
pulse frequency
PW:
pulse width
Vp:
output voltage
Ip:
output current
*Ip:
vp/ 500
Ω
(2) Microcurrent
[f+]=[f-]=f
(pulse frequency)
[PW+]=[PW-]=PW (pulse width)
(500O - 50 k
Ω load)
f=
pulse frequency
PW:
pulse width
Vp:
output voltage
Ip:
output current
*Ip: vp/ 50 k
Ω
1/f
Vp
PW
1/f+
1/f-
PW+
PW-
0
+
-
Vp
Output current:
Constant current, peak value display
Max. 80 mA at 500
Ω load
···TENS, EMS
Max. 750 µA at 50 k
Ω load
···Microcurrent
At maximum output setting
Maximum charge per pulse
Maximum average current
TENS
20 µCoulomb
4 mA
EMS
24 µCoulomb
2.4 mA
MCS
187 µCoulomb
0.3 mA
Pulse width:
50-250 µs
··· TNS
100-300 µs
··· Muscle Stimulation
1-250 ms
··· Microcurrent
Pulse rate:
1-200 Hz
··· TNS Pulse rate and width can be set independently on each channel in Constant mode.
25-100 Hz
··· Muscle Stimulation
0.3-400 Hz
··· Microcurrent
Output mode:
(1)
TNS: Constant, Burst, Modulation, Dual Frequency.
(2)
Muscle Stimulation:
Hold time: 0.5-30 sec.
Rest time: 1-60 sec.
Ramp Up/Down time: 0. 6 sec.
(3)
Microcurrent Constant only
(4)
Program: up to 5 programs
Programming capability:
Number of programs···5
Steps per program·····10
Timer:
30-minute digital timer, with accuracy of +/-2%
Display:
LCD (46 mm x 37 mm)
Other functions: Memory of the previous setting parameters and mode, keyboard lock, automatic power off, low battery power
detection, abnormal pulse output detection and electronics self-test.
Power source:
Dual power source with DC 9 V battery or AC (AC adapter optional)
Classification:
Class II / internally powered equipment, TYPE BF
Conformity:
CE Marking
UL listed:
E203404
Size:
113 mm (H) X 2.7 in (W) X 25 mm (D), (4.4 in (H) X 2.7 in (W) X 1.0 in (D)
Weight:
185 gr. (6.5 oz). Including a battery
Environment for transport and storage:
Temperature:
-10°C - 60°C (14°F - 140°F)
Humidity:
30% - 80%
Specifications and designs are subject to change without prior notice.