Activeforever Mettler Sys*Stim 294 4-Channel Neuromuscular Stimulator User Manual
Sys*stim 294 specifications

1
Sys*Stim 294 Specifications
General Specifications:
Input:
90–240 VAC, 50–60 Hz, 2.3 Amp. Nom.
ETL and C-ETL Listed:
Domestic Model
Model ME 294 (9801427)
Classification:
Protective Class I Equipment
CE model
Type BF Equipment
Enclosed equipment without protection against ingress of
water.
Equipment
not
suitable
for use in the presence of a
flammable anesthetic mixture with air or with nitrogen oxide
Year 2000 Compliant
Yes
Weight: 9.4
pounds
4,3 kg
Dimensions:
5 in (H) x 14.5 in (W) x 10 in (D)
12,7 cm (H) x 36,8 cm (W) x 25,4 cm (D)
Operating Temperature:
+50°F to +104°F
+10°C to +40°C
Humidity:
Operating, 30% to 75% Relative Humidity at 104
°F (40°C)
Nonoperating, 5 to 95% Relative Humidity, non-condensing
Storage Temperature:
-40°F to 167°F
-40°C to 75°C
Storage Humidity:
Storage, 30% to 90% Relative Humidity at 40º C, Non-
condensing
Storage Pressure:
700-900 mB
Environmental Disposal:
The device contains lead in the form of solder used to
produce electrical contact between components. To avoid
adverse environmental impact, utilize a disposal facility that
performs complete incineration of the device at a temperature
in excess of 1000°C.
The shipping materials are fabricated of cardboard and may
be disposed of with other paper products.
Treatment timer:
Timer Accuracy:
±0.5 minutes for times less than 5 minutes
±10% for times from 5 to 10 minutes
±1.0 minute for times greater that 10 minutes
±5%, CE specification
Maximum Treatment Time:
60 minutes–electrical stimulation