Activeforever Mettler Sonicator Plus 940 4-Channel Combination Therapy Device User Manual
Specifications—sonicator® plus 940, General specifications, Ultrasound specifications
Specifications—Sonicator® Plus 940
General Specifications:
Input: 120
V
60 Hz, 85 VA (domestic)
100-240 V
50 Hz, 95 VA (export)
ETL and C-ETL Listed:
Model ME 940
Weight: 11
pounds
Dimensions:
4.9 in (H) x 13.6 in (W) x 10.5 in (D)
Maximum Treatment Time:
60 minutes–electrical stimulation
30 minutes–ultrasound or combination therapy
Treatment Timer:
Treatment time counts down to zero. The digital timer indicates time in
minutes and seconds. The timer also indicates the remaining treatment
time during the “Hold” period.
Ultrasound Specifications:
Frequency: 1.0
MHz
3.0 MHz
Modes: Continuous
Pulsed—5, 10, 20, 30, 40 and 50% duty cycle
Pulse Repetition Rate:
100 Hz
Pulse Duration:
0.5, 1, 2, 3, 4 and 5 msec
Temporal Peak/ average
intensity ratio:
20:1, 10:1, 5:1, 3.3:1, 2.5:1 and 2:1
Maximum output power:
11 W (100%) with large applicator, 1 MHz, (ME 9401)
16.5 W (pulsed) with large applicator, 1 MHz, (ME 9401)
12 W (100%) with large applicator, 3 MHz, (ME 9401)
18 W (pulsed) with large applicator, 3 MHz, (ME 9401)
1.8 W (100%) with small applicator, 1 or 3 MHz, (ME 9402)
2.7 W (pulsed) with small applicator, 1 or 3 MHz, (ME 9402)
Maximum intensity:
2.0 W/cm² (100%)
3.0 W/cm² (pulsed mode)
Indication accuracy:
±20% (for any level above 10% of maximum)
Individual Applicator Specifications:
Applicator Part
Number
Frequency
Effective Radiating Area
Maximum Beam
Non-Uniformity Ratio
Type
ME 9401
1 MHz
5.5 cm²
4.6:1
Collimated
ME 9401
3 MHz
6.0 cm²
4.2:1
Collimated
ME 9402
1 MHz
0.9 cm²
4.7:1
Divergent
ME 9402
3 MHz
0.9 cm²
4.7:1
Collimated
Electrical Stimulation Specifications:
4-Pole Interferential Mode
4-Pole Interferential Waveform
Waveform Type:
Sinewave
Polarity:
None
Maximum Voltage:
50 V (Peak value, 500Ω load)
Current:
0–100
mA
peak,
500Ω
load
Carrier Frequency:
2 kHz, 4 kHz, 5 kHz
Interference frequency
1~250 beats
Frequency Modulation:
1 to 10 in 1-beat steps, and
10 to 250 in 10-beat steps.
Vector sweep angle:
0, 15, 30, 45 degrees
Available Channels:
Channels 1 & 2 or 3 &4