TYPICAL PRODUCT PARAMETERS
Part No 1148139
5
MA90 and MA95 Series Mattresses
TYPICAL PRODUCT PARAMETERS
NOTE: MA90Z power unit is also used on MA90ZB42 and MA90ZB48 systems.
NOTE: MA95Z power unit is also used on MA95ZB42 and MA95ZB48 systems.
Electrical Parameters
Performance Parameters
MA90Z
MA95Z
INPUT VOLTAGE AC:
110 V
INPUT FREQUENCY:
60 Hz
CURRENT:
1 A
2 A
MAXIMUM POWER CONSUMPTION:
30 ± 10 W
150 ± 10 W
CIRCUIT PROTECTION:
Dual fused,
250 V,
1 A fast blow fuses
Dual fused,
250 V,
5 A fast blow fuses
MODE OF OPERATION:
Continuous
MA90Z
MA95Z
WEIGHT CAPACITY
STANDARD MATTRESS:
BARIATRIC MATTRESS
42 inch:
48 inch:
350 lbs
650 lbs
1000 lbs
PRESSURE ZONE:
2
MAXIMUM FLOW:
50 ± 15 LPM
1275 ± 100 LPM
MAXIMUM PRESSURE:
35 ± 5 mmHg
MAXIMUM FLOW TIMER:
30 minutes
15 minutes
SUPPORT SURFACE INFLATION TIME:
5 - 10 minutes
25 - 60 seconds
PATIENT COMFORT CONTROL PRESSURES:
SOFT PRESSURE:
FIRM PRESSURE:
8 ± 4 mmHg
32 ± 4 mmHg
ROTATION TIME:
10, 20, 30, 60 minutes
5 minutes - 4 hrs
ROTATION ANGLE:
0 - 30° ± 5°
0 - 40° ± 5°
PATIENT CONTACT:
Control unit and mattress have
Latex-Free components