Location recommendations for optimal performance, Location, Recommendations for optimal performance – Activeforever Invacare XPO2 Portable Oxygen Concentrator User Manual
Page 18: Ƽ warning

SECTION 6—OPERATING INSTRUCTIONS
18
Part No 1148112
Location
ƽ WARNING
NEVER block the air openings of the product or place it on a
soft surface, such as a bed or couch, where the air opening
may be blocked. Keep the openings free from lint, hair and the
like.
Keep unit at least three inches away from walls, draperies,
furniture, and the like.
Locate and position the XPO
2
in a well ventilated space so that
the air intake and the air exhausts are not obstructed.
Recommendations for Optimal
Performance
Operating Temperature:
41°F to 95°F (5°C to 35°C)
Relative Humidity:
15% to 60%
Transport/Storage
Temperature:
-2°F to 140°F (-20°C to 60°C)
Allow unit to warm, or cool, to operating
temperature range before using.
Electrical:
No extension cords.
Altitude:
Up to 10,000 ft (3046 m) above sea level.
Tubing and Cannula:
4 ft cannula with a maximum 25 ft of crush
resistant tubing (DO NOT pinch).
Environment:
Smoke, pollutant and fume free. No
confined spaces (example: no closets).
Time of Operation:
Up to 24 hours per day when connected to
AC or DC power adapter.