Display.messages.and.troubleshooting, Warning – Accu-Chek Aviva Plus silver User Manual
Page 84
82
Battery power is low. Change the
battery soon. See Chapter 6,
“Changing the Battery.”
The meter is in the set‑up mode,
waiting for you to change or confirm
settings. See Chapter 4, “Meter
Setup.”
Display.Messages.and.Troubleshooting
WARNING
Never.make.treatment.decisions.based.on.an.error.message..If.you.have.any.concerns,.contact.the.
ACCU‑CHEK.Customer.Care.Service.Center.at.1‑800‑858‑8072.
The meter will not turn on or the
display is blank.
• Battery is dead –
Insert new battery.
• Display is damaged –
Contact the ACCU‑CHEK Customer
Care Service Center at
1‑800‑858‑8072.
• Extreme temperatures –
Move the meter to a more
temperate area.
(blank.display)