5. Data Memory

At the end of each measurement, this instrument automatically

stores the result.
Viewing the stored values

Press the M-button AO briefly, when the instrument is switched off.

The display first shows the last stored result.

Pressing the M-button again displays the previous value. Pressing

the M-button repeatedly enables you to move from one stored

value to another.
Memory full

ƽ

When the memory has stored 30 results, the memory is full.

From this point onwards, a new measured value is stored

by overwriting the oldest value.

Clearing all values

ƽ

All memory data are cleared when batteries are taken out

of the instrument.

6. Battery Indicator and Battery change

Low battery

When the batteries are approximately ¾ empty the battery symbol
AK will flash as soon as the instrument is switched on (partly filled

battery displayed). Although the instrument will continue to

measure reliably, you should obtain replacement batteries.
Flat battery – replacement

When the batteries are flat, the battery symbol AK will flash as soon

as the instrument is switched on (flat battery displayed). You cannot

take any further measurements and must replace the batteries.

1. Open the battery compartment 4 at the back of the instrument.

2. Replace the batteries – ensure correct polarity as shown by the

symbols in the compartment.

Which batteries and which procedure?

Use 4 new, long-life 1.5V, size AA batteries.

Do not use batteries beyond their date of expiry.

Remove batteries if the instrument is not going to be used

for a prolonged period.

Using rechargeable batteries

You can also operate this instrument using rechargeable batteries.

Only use «NiMH» type reusable batteries.

Batteries must be removed and recharged when the flat

battery symbol appears. They should not remain inside the

instrument as they may become damaged (total discharge as

a result of low use of the instrument, even when switched off).

Always remove the rechargeable batteries if you do not

intend to use the instrument for a week or more.

Batteries cannot be charged in the blood pressure monitor.

Recharge batteries in an external charger and observe the

information regarding charging, care and durability.

7. Error Messages

If an error occurs during the measurement, the measurement is

interrupted and an error message, e.g. «ERR 3», is displayed.

* Please consult your doctor, if this or any other problem occurs

repeatedly.

If you think the results are unusual, please read through the

information in «Section 1.» carefully.

Error

Description Potential cause and remedy

«ERR 1» Signal too

weak

The pulse signals on the cuff are too weak.

Re-position the cuff and repeat the

measurement.*

«ERR 2» Error signal During the measurement, error signals were

detected by the cuff, caused for instance by

movement or muscle tension. Repeat the

measurement, keeping your arm still.

«ERR 3» No pressure

in the cuff

An adequate pressure cannot be generated

in the cuff. A leak may have occurred.

Check that the cuff is correctly connected

and is not too loose. Replace the batteries if

necessary. Repeat the measurement.

«ERR 5» Abnormal

result

The measuring signals are inaccurate and

no result can therefore be displayed. Read

through the checklist for performing reliable

measurements and then repeat the

measurement.*

«HI»

Pulse or

cuff pres-

sure too

high

The pressure in the cuff is too high (over

300 mmHg) OR the pulse is too high (over

200 beats per minute). Relax for 5 minutes

and repeat the measurement.*

«LO»

Pulse too

low

The pulse is too low (less than 40 beats per

minute). Repeat the measurement.*