Microlife BP 2BHO User Manual
Page 14
![background image](/manuals/234786/14/background.png)
9. Technical specifications
Weight:
575 g (with batteries + cuff size M)
Size:
106 (W) x 162 (L) x 61.5 (H) mm
Storage temperature:
–5 to +50 °C
Humidity:
15 to 90% relative humidity maximum
Operation temperature:
10 to 40 °C
Display:
LCD-Display (Liquid Crystal Display)
Measuring method:
oscillometric
Pressure sensor:
capacitive
Measuring range:
SYS/DIA:
30 to 280 mmHg
Pulse:
40 to 200 per minute
Cuff pressure display range:
0–299 mmHg
Memory:
Storing the last measurements automatically
Measuring resolution:
1 mmHg
Accuracy:
Pressure:
within ± 3 mmHg
Pulse:
± 5 % of the reading
Power source:
a) 4 x 1.5 V Batteries; size AA
b) Mains adapter DC 6V, 600 mA (optional)
Cuffs:
type M for arm circumference 22–32 cm or
type L for arm circumference 32–42 cm (optional)
Reference to standards:
EU Directives 93/42/EEC; NIBP requirements:
EN 1060-1 /-3 /-4, ANSI / AAMI SP10
Technical alterations reserved!
10. www.microlife.com
Detailed user information about our thermometers and blood pressure monitors as well as services can be
found at www.microlife.com.
11