Parameter lists 4-5 – VXI VT1433B User Manual
Page 83

Parameter Lists
This section shows which parameters are global parameters, which are
channel-specific and what types of channels the channel-specific parameters apply
to. Default values are shown for all of these parameters. In addition, each
parameter is categorized as “abort”, “wait”, “immediate” or “glitch” depending on
the behavior when this parameter is changed during a running measurement.
Those with “abort” cause the measurement to abort. Those with “wait” don’t take
effect until the start of the next measurement. Those with “immediate” take
effect immediately. Those with “glitch” take effect immediately and may cause
glitches in the data that is read back or on the source output if the parameter is
applied to a source channel.
Global Parameters
Parameter
Default Value
Changes
append_status
Off
Immediate
arm_channel
None
Immediate
arm_mode
Auto Arm
Immediate
arm_time_interval
1 s
Immediate
auto_group_meas
On
Wait
avg_mode
None
Wait
avg_number
10
Wait
auto_trigger
Auto Trigger
Abort
avg_update
10
Wait
avg_weight
1
Immediate
blocksize
1024
Abort
cal_dac
0
Immediate
cal_voltage
0 V
Immediate
calin
Grounded
Immediate
center_freq
2 kHz
Immediate
clock_freq
51.2 kHz
Abort
clock_master
Off
Abort
clock_source
Internal
Abort
data_mode
Block Mode
Abort
data_port
VME
Abort
data_size
16 Bit Integer
Abort
VT1433B User's Guide
The Host Interface Library