Table of Baseband Measurement Spans (part 6 of 6)
All values are in Hertz (Hz).

sample frequency

(Fs) —>

163840

192000

196608

number of

÷2 steps

with ÷5

without ÷5

with ÷5

without ÷5

with ÷5

without ÷5

0

12800.000000

64000.000000*

15000.000000

75000.000000*

15360.000000

76800.000000*

1

6400.000000

32000.000000*

7500.000000

37500.000000

7680.000000

38400.000000

2

3200.000000

16000.000000

3750.000000

18750.000000

3840.000000

19200.000000

3

1600.000000

8000.000000

1875.000000

9375.000000

1920.000000

9600.000000

4

800.000000

4000.000000

937.500000

4687.500000

960.000000

4800.000000

5

200.000000

2000.000000

468.750000

2343.750000

480.000000

2400.000000

6

200.000000

1000.000000

234.375000

1171.875000

240.000000

1200.000000

7

100.000000

500.000000

117.187500

585.937500

120.000000

600.000000

8

50.000000

250.000000

58.593750

292.968750

60.000000

300.000000

9

25.000000

125.000000

29.296875

146.484375

30.000000

150.000000

10

12.500000

65.500000

14.648438

73.242188

15.000000

75.000000

11

6.250000

31.250000

7.324219

36.621094

7.500000

37.500000

12

3.125000

15.625000

3.662109

18.310547

3.750000

18.750000

13

1.562500

7.812500

1.831055

9.155273

1.875000

9.375000

14

0.781250

3.906250

0.915527

4.577637

0.937500

4.687500

15

0.390625

1.953125

0.457764

2.88818

0.468750

2.343750

16

0.195312

0.976562

0.228882

1.144409

0.234375

1.171875

Notes:

* For the top span, the 3 dB bandwidth is 1.15 times span shown.

To select sample frequency for time domain data, first divide the desired sample
frequency by 2.56 to convert it to a measurement span. Then locate the closest
measurement span on this table and choose the corresponding sample frequency at
the top of the table.

VT1433B User's Guide

Using the VT1433B

3-23