Caution, 4 grounding – Toshiba GF632 User Manual
Page 26

 
 
- 25 -
6F8A0883
4.4 Grounding
CAUTION
 Do not wire cables and replace parts
when power is supplied.
 Do not work on piping and wiring with wet hands.
DON’T
Wiring work and replacing 
parts in the power-on state 
may cause electric shock. 
DON’T
Wet hands may result in electric shock.
(1) Grounding of the GF630/LF600F and GF630/LF610F type
Ground as shown in Figure 4.12. Make the grounding wire as short as possible. Use grounding 
wire material of IV wire 5.5mm
2
or more. Do not share a grounding wire with other
instruments where grounding current may flow. (An independent grounding is preferable.)
 
 
 
 
 
 
Figure 4.12 Grounding the GF630/LF600F and GF630/LF610F Type
 
Please connect the ground cables 
of detector to piping using the M5 screw 
shown as Figure4.13 as needed. 
(In the case that there are no threaded hole 
with piping, please tap the threaded hole.)
※It is necessary to connect the ground cable 
 surely electrically. 
 
 
grounding resistance 
100ohm or less 
•In the case that the piping is conductive material, 
connect the ground cables to the both ends of the 
piping flange. 
• In the case that the piping is non-conductive material, 
use grounding rings. 
Integral
Conductive material pipe 
Example: Metal, etc. 
Piping of non-conductive material 
Example: Resin or metal piping whose 
inside is resin-lined, etc. 
Grounding wire
grounding ring
Grounding terminal
grounding resistance 
100ohm or less 
A
Figure4.13 Grounding wire connection
(Detail of part A)
M5
Ground cable of detector