Quickstart Guide

Trace Oxygen Analyzer


Teledyne Analytical Instruments

4

SETUP AND INSTALL

Mount the Analyzer

The Model 3000TA -EU is for indoor use in a general purpose area. It is NOT for hazardous

environments of any type. It is designed for flush panel mounting. Use the four mounting holes—
one in each corner of the rigid frame to panel mount the instrument.

Make Gas Connections

Note: Do not remove plastic caps on VCR fittings, unless you are ready to make

connections.

The analyzer is equipped with 1/4

inch tube fittings, and 6 mm adapters are supplied for

metric system installations.

1. Connect your sample gas to the

SAMPLE IN

port.

2. Connect a vent line to

EXHAUST OUT

port.

3. Calibration gases should be tee'd into the sample inlet with appropriate valves.
4. Regulate the sample gas pressure between 2 and 50 psig sufficient to keep the front

panel flowmeter reading in an acceptable range (0.5 to 2.0 SLPM). For non-
pressurized sample or very low pressure, (less than 2 psig) vacuum service plumbing
is recommended.

5. If greater sample flow is required for improved response time, install a bypass in the

sampling system upstream of the analyzer input.

6. For optional vacuum application where the sample pressure is at atmospheric or very

low pressure, the instrument should be fitted with the vacuum service option. This
option locates the flow control valve on the exhaust side of the Micro-fuel Cell.

Note: Exhaust connections must be consistent with the hazard level of the constituent

gases. Check Local, State, and Federal laws, and ensure that the exhaust
stream vents to an appropriately controlled area, if required.