Model 3000 TA-XL-EU

Quickstart Guide

Teledyne Analytical Instruments

13

#### ppm Span

Slope=#### ppm/s

The instrument automatically returns to the Analyze mode at the conclusion of the auto span

cycle..

Manual Mode Spanning:

Press Span to start the Span function.

Span: Settling:MAN

For Next

Use the ▲▼ keys to toggle between

AUTO

and

MAN

span settling. Stop when

MAN

appears, blinking, on the display. Press Enter to move to the next screen.

Calib. Holding time

Cal hold: 5 min

Press Enter key to continue.

Span Val: 000008.00

Span Mod #

Press ▲ () to permit modification (

Mod #

) of span value.

Use the arrow keys to enter the oxygen concentration of the span gas you are using
(209000.00 if you are using air). The ◄► arrows choose the digit, and the ▲▼
arrows choose the value of the digit.

Press Enter to enter the span value into the system and begin the span calibration.

#### % Span

Slope=#### ppm/s

When the Span value displayed on the screen is sufficiently stable, press Enter and
the span reading changes to the correct value.

The instrument automatically returns to the Analyze mode.

Setting Alarm Parameters

If you are using password protection, you will need to enter your password to access the

alarm functions. Follow the instructions above for Entering the Password to enter your
password. Once you have clearance to proceed, you can set the alarms.