Model 3000 TA-XL-EU

Quickstart Guide

Teledyne Analytical Instruments

11

◄► arrow keys to scroll back and forth between letters, and the ▲▼ arrow keys to
change the letters to the proper password.

Press Enter to enter the password.

Change Password?

=Yes =No

Press Escape.

Calibration of the Analyzer

For trace analysis, the analyzer must be purged before calibration. Refer to the Instruction

manual for purging procedure.

Zero Calibration

Make sure an appropriate zero gas is connected to the sample port for this procedure.


Auto Mode Zero Calibration

Press Zero to enter the zero function mode.

Use the ▲▼ arrow keys to toggle between

AUTO

and

MAN

zero settling. Stop when

AUTO

appears blinking.

Zero: Settling: AUTO

To Begin

Press Enter to begin zeroing.

#### PPM Zero

Slope=#### ppm/s

.
.
.

#### PPM Zero

4 Left=### ppm/s