Basic operation – Teledyne Trace Oxygen Analyzer 3000TA-XL-EU User Manual
Page 10
Quickstart Guide
Trace Oxygen Analyzer
Teledyne Analytical Instruments
8
BASIC OPERATION
Once the analyzer has been installed, it can be configured for your application. Prior to
using the instrument you must:
•
Calibrate the instrument
•
Define the analysis ranges and choose whether to use autoranging or a fixed range.
•
Set alarm setpoints and modes of alarm operation.
Additional configuration parameters are available to tailor the instrument to your particular
application. Refer to the Instruction Manual for specific information on setting of additional
system parameters.
Default Configuration
The default analyzer configuration is as follows:
Ranges: LO = 1 ppm, MED = 10 ppm, HI = 100 ppm.
Auto Ranging: ON
Alarm Relays: Defeated, 10.00 ppm, HI, Not failsafe, Not latching.
Zero: Auto, every 0 days at 0 hours.
Span: Auto, at 000,008.00 ppm, every 0 days at 0 hours.
If you choose not to use password protection, the default password is automatically
displayed on the password screen when you start up, and you simply press Enter for access to all
functions of the analyzer.
Using the Data Entry and Function Buttons
The ◄► arrow buttons select options from the menu currently being displayed on the
display. When selected, the option blinks.
When the selected option includes a modifiable item, the ▲▼ arrow buttons can be used to
increment or decrement that modifiable item.
The Enter button is used to accept any new entries on the VFD screen.
The Escape button is used to abort any new entries on the VFD screen that are not yet
accepted by use of the Enter button.