Teledyne Trace Oxygen Analyzer 3000TA-XL-EU User Manual

Page 14

background image

Quickstart Guide

Trace Oxygen Analyzer


Teledyne Analytical Instruments

12

The zeroing process will automatically conclude when the output is within the acceptable

range for a good zero. Then the analyzer automatically returns to the Analyze mode.

Manual Mode Zeroing

Press Zero to enter the Zero function.

Use the ▲▼ keys to toggle between

AUTO

and

MAN

zero settling. Stop when

MAN

appears, blinking, on the display.

Zero: Settling: Man

To Begin

Press Enter to begin the zero calibration.

#### ppm Zero

Slope=#### ppm/s

Once zero settling is completed, the instrument automatically returns to the Analyze mode.

Span Calibration

Make sure an appropriate span gas is connected to the sample port.


Auto Mode Spanning:

Press Span to enter the span function.

Use the arrow keys to toggle between

AUTO

and

MAN

span settling. Stop when

AUTO

appears, blinking on the display.

Span: Settling: AUTO

For Next

Press Enter.

Calib. Holding time

Cal hold: 5 min

Press Enter to move to the next screen.

Span Val: 000008.00

Span Mod #

Use the ▲▼ arrow keys to enter the oxygen-concentration mode. The ◄► arrows
choose the digit, and the ▲▼ arrows choose the value of the digit. When you have
finished typing in the concentration of the span gas you are using, press Enter to
begin the span calibration.