Teledyne Trace Oxygen Analyzer 3000TA-XL-EU User Manual
Page 14
![background image](/manuals/209673/14/background.png)
Quickstart Guide
Trace Oxygen Analyzer
Teledyne Analytical Instruments
12
The zeroing process will automatically conclude when the output is within the acceptable
range for a good zero. Then the analyzer automatically returns to the Analyze mode.
Manual Mode Zeroing
•
Press Zero to enter the Zero function.
•
Use the ▲▼ keys to toggle between
AUTO
and
MAN
zero settling. Stop when
MAN
appears, blinking, on the display.
Zero: Settling: Man
•
Press Enter to begin the zero calibration.
#### ppm Zero
Slope=#### ppm/s
Once zero settling is completed, the instrument automatically returns to the Analyze mode.
Span Calibration
Make sure an appropriate span gas is connected to the sample port.
Auto Mode Spanning:
•
Press Span to enter the span function.
•
Use the arrow keys to toggle between
AUTO
and
MAN
span settling. Stop when
AUTO
appears, blinking on the display.
Span: Settling: AUTO
•
Press Enter.
Calib. Holding time
Cal hold: 5 min
•
Press Enter to move to the next screen.
Span Val: 000008.00
•
Use the ▲▼ arrow keys to enter the oxygen-concentration mode. The ◄► arrows
choose the digit, and the ▲▼ arrows choose the value of the digit. When you have
finished typing in the concentration of the span gas you are using, press Enter to
begin the span calibration.