7.6 Status Output Summary Table

Table 7-2 below is a logic truth table summarizing the recommended actions based on the
states of the four status outputs. A ‘1’ indicates the output is ON, a ‘0’ indicates the
output is OFF, and ‘X’ indicates the output is in either state.

Table 7-2: Status Output Truth Table

Sensor
OK

Invalid
Reading

Lamp
Low

Cell
Dirty

Actions

1

0

0

0

Normal operation, no action
required

0 X X

X

Service

required

1

1

X

X

Check Pressure > 45 psia
Verify that concentration has
not exceeded full scale range
of sensor.
Calibrate at Zero.

1

X

1

X

Lamp adjustment useful,
though not required

1

X

X

1

Calibrate at zero
Clean Cell