7.6 Status Output Summary Table
Table 7-2 below is a logic truth table summarizing the recommended actions based on the
states of the four status outputs. A ‘1’ indicates the output is ON, a ‘0’ indicates the
output is OFF, and ‘X’ indicates the output is in either state.
Table 7-2: Status Output Truth Table
Sensor
OK
Invalid
Reading
Lamp
Low
Cell
Dirty
Actions
1
0
0
0
Normal operation, no action
required
0 X X
X
Service
required
1
1
X
X
Check Pressure > 45 psia
Verify that concentration has
not exceeded full scale range
of sensor.
Calibrate at Zero.
1
X
1
X
Lamp adjustment useful,
though not required
1
X
X
1
Calibrate at zero
Clean Cell