1.4 Specifications

Measurement Principle

UV Absorption (Beer Lambert Law)

Ranges

0-5%, 0-10%, 0-15% w/w, 0-20% w/w
0-100 g/Nm

3

, 0-200 g/Nm

3

, 0-300 g/Nm

3

, 0-

400g/ Nm

3

Measurement Units

wt%, g/Nm

3

Accuracy

± 1% of Full Scale

Precision/Repeatability

±0.5% of Full Scale Range

Resolution

0.1 % w/w or .01 g/Nm

3

Response Time (95%)

<5 sec to 95%

Compensation

Pressure, Temperature (NTP = 273.15K,
760mmHg)

Gas Inlet Pressure Range

3.0 – 30.0 psig

Gas Flow Rate

0.2 – 2.0 LPM

Temperature Range

5-45oC

Dimensions (H x W x D)

12.64" x 11.19" x 6.08"
(321mm x 284mm x 154mm)

Weight 9.40lb

(4.27kg)

Power

110-240V~, 50/60Hz, 2.5A

Environmental Conditions

Installation Category (Overvoltage Category) II
Pollution Degree 2

Maximum Operating Altitude

2000 meters

Analog Output Voltage

0-5V, 4-20mA (Optional)

Isolated Analog Output 4-20mA
Mode*

Maximum voltage between outputs and ground
60V peak

Degree of Protection (IP Code)

IPX65

*Optional











P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 10