5.4.2 CZERO
Syntax
<address>CZERO<CR>
Description
Perform a Zero Calibration. Note that calibration is performed based on concentration
currently in measurement cell. Instrument should be sufficiently purged with zero gas
before issuing this command.
Data Parameters Sent
None
Response
<address>:<success_flag>#<checksum><CR>
Example
Command:
1CZERO
Response:
1:OK#261
- Calibration Successful
1:FAIL#391
- Calibration Failed