6 tdump, Syntax, Description – Teledyne 460H User Manual
Page 28: Data parameters sent, Response, Example, Notes
5.4.6 TDUMP
Syntax
<address>TDUMP<CR>
Description
Returns the current measured parameters: O3 Concentration, Cell Pressure(psia,) Cell
Temperature (K,) Lamp Temperature (K,) Measure Detector (mV,) Calibrated
Refererence Detector (mV,) Reference Detector (mV)
Data Parameters Sent
None
Response
<address>:<o3_conc>,
Example
Command:
1TDUMP
Response:
1:0.0282144,14.77461,300.7179,324.7713,2881.437,2940.903,4412.52#3228
Notes
Note that the concentration value returned shows more digits after the decimal than the
front panel display. Concentration values should only be considered valid to 4 significant
digits.
P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 28