beautypg.com

8 vget, Syntax, Description – Teledyne 460H User Manual

Page 30: Data parameters sent, Response, Example

background image

5.4.8 VGET

Syntax

VGET:<var_index>

Description
Returns value of an internal configuration variable (VAR.)

Data Parameters Sent
var_index

is index number for internal VAR as follows:

Var

Index

Name

Description

Allowable Range

0 analog_range

Full-Scale

concentration

range for Analog Output
scaling

5.0 – 400.0 (Units are
current unit of measure,
wt% or g/Nm3)

1

azero_enable

Auto-Zero Enable VAR.
(only valid if Auto-Zero
hardware is installed)

0 = Off
1 = On

2

azero_period

Period for Auto-Zero
Calibration timer

5.0 - 86400.0 Seconds

3

carrier_weight

Mol weight of carrier gas
for wt% calculations (32.0
= O2)

27.0 – 32.0 mol weight

4 comm_mode

Not

Used

N/A

5 iir_filt

Digital

Concentration

filter

0.05 – 1.0
(1.0 = No Filtering, lower
values increases filtering)

6 conc_units

O3

Concentration

measurement units

0 = wt%
1 = g/Nm3


Response
<address>:<var_value>#<checksum><CR>

Example
Command:

1VGET:0<CR>

- Request analog_range value


Response:

1:15.0#303

- analog_range returned as 15.0 (wt%)

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 30