Omega CL27 User Manual

Page 33

background image

31

Current

Sensor

Unit

Source

DMM

Step

Select

Output

Input

Adjust

Verify

Reading*

1

RTD-100

1562.0°F

1mA

P38

0.39026v

2

RTD-100

32.0°F

1mA

0.10000

±0.00003v

3

RTD-100

–328.0°F

1mA

0.01849

±0.00006v

4

RTD-1000

1562.0°F

100µA

P51

0.39026v

5

RTD-1000

32.0°F

100µA

0.10000

±0.00003v

6

1,000

900.00

1mA

0.90000

±0.00010v

7

1,000

1.00

1mA

0.00100

±0.00010v

8

10,000

9,000.0

100µA

0.90000

±0.00010v

9

10,000

10.0

100µA

0.00100

±0.00010v

Table 4: Calibration of Calibrator-Mode Resistance Functions

(RTD-100/1000)

* exclusive of noise.

Sensor

Unit

Unit

Step

Select

Input

Adjust

Verify

Reading*

1

RTD-100

390.26

P1

1562.0°F

2

RTD-100

100.00

32.0±0.3°F

3

RTD-100

18.49

–328.0±0.5°F

4

1,000

900.00

P1

900.00

5

1,000

1.00

1.00±0.20

6

100,000

90,000

P1

90,000

7

100,000

100

100±20

8

Thermistor

47.0

P1

293.0°F

9

Thermistor

2252

77.0±0.2°F

10

Thermistor

75,790

–40.0±0.2°F

Table 5: Calibration of Meter-Mode Resistance Functions

(RTD-Thermistor)

* exclusive of noise.