1. DIAGNOSTICS AND ERROR MESSAGES

Unit self-diagnostics provide fault condition readouts which are described
below:

PROBLEM:

LIKELY CAUSE:

Blank display, unit does not

(1) Improper battery installation.

power-up.

(1)

Check battery polarity.

(2) Dead battery.

BAT Annunciator.

(1) Low battery voltage, install a new

battery. If problem persists, con-
sult factory.

Display reads OPEN.

(1) No thermocouple, RTD or thermis-

tor is plugged into the meter input.

Display reads E1 momentarily.

(1) This indicates that an invalid key-

pad entry has been made. Review
keystroke sequence, or consult
manual for input instructions.

Display reads E2 momentarily.

(1) Attempt made to activate non-

installed ramp function. NOTE:
RAMP Function enabled on
Models CL25 and CL27 only.

Display reads HI.

(1) Meter-mode input-temperature

exceeds rating of selected sensor.

(2) Calibrator-mode keypad-entry

exceeds rating of sensor type.

Display reads LO.

(1) Meter-mode input-temperature too

low for accurate measurement.

(2) Calibrator-mode keypad-entry too

low for accurate simulation.

2. INITIAL TURN ON

Turn on the instrument by depressing the ON-OFF keyswitch once (Figure
2). At power-up, the unit first performs a display test. All segments and
annunciators of the liquid crystal (LCD) are momentarily turned on for visual
confirmation by the user (Figure 3).

NOTE: In case of missing or poor-contrast segments, contact

factory.