beautypg.com

Omega CL27 User Manual

Page 32

background image

Sensor

Unit

Unit

Step

Select

Input

Adjust

Verify

Reading*

1

RTD-100

390.26

P1**

1562.0°F±0.5°F

2

RTD-100

100.00

32.0±0.1°F

3

RTD-100

18.49

–328.0±0.2°F

4

RTD-1000

3,902.6

P1**

1562.0°F±0.5°F

5

RTD-1000

1000.0

32.0+0.1°F

6

1,000

900.00

P1

900.00

7

1,000

1.00

1.00±0.10

8

10,000

9,000.0

P1

9000.0

9

10,000

10.0

10.0±1.0

Table 3: Calibration of Meter-Mode Resistance Functions

(RTD-100/1000)

* exclusive of noise.
** adjust as necessary to obtain steps 2, 5.

Calibration Adjustments:

NOTE: Do not deviate from the calibration adjustment sequence
that follows. This will ensure that adjustments to be stored in
EEPROM go to the correct locations.

1. Perform meter-mode adjustments and verifications per Table 3 (use Table

5 for RTD-Thermistor models).

2. Hit OPR key (Flashing 'RTD' annunciator turns off).

3. Install J1 jumper.

4. Set unit to CALIB mode, RTD-100, °F.

5. Switch current source to OPERATE mode.

6 Perform calibrator-mode adjustments and verifications per Table 4 (use

Table 6 for RTD-Thermistor models).

7. Turn off unit. Remove tape holding battery.

8. Re-install back cover. Resistance-function calibration is complete.

30

This manual is related to the following products: