beautypg.com

Omega CL27 User Manual

Page 30

background image

B. RESISTANCE CALIBRATION

NOTE: Thermocouple and resistance calibrations are indepen-
dent. However, potentiometer P1 is used in both thermocouple
and resistance calibrations to set EEPROM constants. Follow cal-
ibration procedures carefully to ensure that thermocouple and
resistance calibrations do not disturb one another.

Test Equipment Required:

1. Resistance Decades (General Resistance RTD-100, RTD-500/1000, and

RDS-54, or equivalent).

2. Current Source (Fluke 5101B, or equivalent).

3. Digital Multimeter (OMEGA OM7551, or equivalent).

4. Calibration cables per Figure 14.

Ambient Conditions:

Unit should be calibrated at an ambient temperature of 23°C±1°C, with rela-
tive humidity less than 80%.

Preparation For Calibration:

1. Turn off unit, and remove bottom-cover. Refer to Disassembly

Instructions.

2. Secure battery with a piece of adhesive tape.

NOTE: Momentary loss of battery power during the following cal-
ibration will invalidate previous adjustments, both resistance and
thermocouple.

3. Hook up test-equipment and unit per Figure 14.

4. Set current-source to 1mA, STANDBY mode.

5. Set DMM to 2V DC range.

6. Turn on unit.

7. Set unit to METER mode, RTD-100, °F.

8. Turn off unit. Remove jumper J1. Turn on unit. 'RTD' annunciator should

be flashing.

NOTES: Flashing 'RTD' indicates that calibration adjustments

will be written to memory. Avoid selecting thermocouple
functions while in this mode.

NOTES:

Should it be necessary to scroll through thermocouple
functions, first hit the OPR key (disables flashing 'RTD'),
and start again at Step 7 above.

28

This manual is related to the following products: