SERVICE INFORMATION

WARNING

All service information is intended for qualified electronic mainte-
nance personnel only.

1. CALIBRATION PROCEDURES

This is a two-part procedure. Products with thermocouple functions only, fol-
low the Thermocouple Calibration procedure below. Products with resistance
functions (RTD, thermistor, ohms) follow first the Thermocouple Calibration
procedure, then the Resistance Calibration procedure.

A. THERMOCOUPLE CALIBRATION

Test Equipment Required:

1. Thermocouple simulator (OMEGA CL521, or equivalent) calibrated to
ITS-90.

2. Microvolt DMM (OMEGA OM7551, or equivalent)

3. Calibration cables, per Figures 11, 12, 13.

NOTE: The copper constantan cables used in Figures 12 and 13
need to be calibrated. Voltage errors in these cables should be com-
pensated at the simulator.

4. Calibration cover (OMEGA 820-307-4).

Ambient Conditions:

Units should be calibrated at an ambient temperature of 23°C ±1°C, with rel-
ative humidity less than 80%.

Preparation for Calibration:

1

1. Remove battery from unit under test (U.U.T.). Refer to Battery

Installation/Replacement section in the manual.

1

2. Remove bottom-cover from U.U.T. Refer to Disassembly Instructions.

1

3. Remove calibration-jumpers J1 and J2 from printed-circuit board.

NOTE: Removal of J1, J2 causes partial loss of previously stored
calibration data. Do not remove these jumpers unless recalibra-
tion is intended.