Maintenance and inspection, Troubleshooting – Omron Healthcare INTELLISENSE HEM-907XL User Manual

Page 6

background image

6

NOTES ON SAFETY

Maintenance and inspection

1. Check the unit operation on regular basis.

2. If this unit has not been used for more than three months, be sure to check that this unit

operates normally and safely before use.

Troubleshooting

If device error 9 (Er9) occurs, take the following procedure promptly:

(1) Remove the cuff from the patient’s arm.

(2) Turn off the power of the unit and unplug the AC adapter from the electric outlet.

(3) Display “Out of use” on this unit so that it cannot be used.

(4) Contact Omron Healthcare’s Customer Service at 1-877-216-1336.